Digital Dissolved Oxygen Meter Do Meter Industrial Sewage Aquaculture Dissolved oxygen saturation tester Analyzer 0.0-40.0mg/l
Description:
Calibration method: calibrate in air.
In order to facilitate transportation, the electrode filling solution is replaced by powder.
Preparation of electrode filling solution:
1 pack of electrode filling solution powder with 500ml purified water, add 1.5ml electrode film sleeve when using, and store at low temperature after using.
Measuring range
Oxygen concentration (mg / L): 0.0-40.0mg/l
Dissolved oxygen saturation (%): 0.0-300.0%
Temperature: 0.1-40.0 ℃; 32.0-104.0F
resolving power
Oxygen concentration (mg / L): 0.1mg/l
Dissolved oxygen saturation (%): 0.1%
Temperature: 0,1 ℃; 0.1F
Accuracy:
Oxygen concentration (mg / L): ± 1.5% F.S
Dissolved oxygen saturation (%): ± 1.5% F.S
Temperature: ± 0.5 ℃
Support single point or two-point automatic calibration, i.e. zero point or in air
Calibration mode: automatic 100% calibration
Response time: ≤ 45s (fast response at 20 ℃)
Temperature compensation range: automatic (0-40 ℃)
Height compensation: None
Battery power: low power warning
Backlight: support automatic backlight
Shutdown: automatic shutdown for about 8 minutes
Power supply mode: 3 * 1.5V button battery (not include)
Service environment: 0.1-40.0 ℃ RH max 90%
Size: 183 * 37 * 37mm (7.2 * 1.5 * 1.5in)
Weight: 80g
Do calibration procedure
The calibration procedure has one-point calibration or two-point calibration
One point calibration: (to meet normal use)
0.0% (oxygen free water) or 100.0% (air).
Two point calibration:
Simultaneous calibration of 0.0% (oxygen free water) and 100.0% (air)
The calibration of the instrument is simple. Make sure that the probe is installed correctly before calibration
100.0% - calibrate in the air after the value is stable in the power on state. Long press the cal key for 5 seconds and release the key. The screen flashes 100.0%. After flashing, the calibration is completed.
0.0% - calibrate in anaerobic water when the value is stable in the power on state. Long press the cal key for 5 seconds and release the key. The screen flashes 0.0%. After flashing, the calibration is completed.