VC77 Digital Multimeter 2 In 1 4-20MA Signal Output Multimeter Process Signal Source
Tax included.
1.Multimeter & Signal Source 2 in1
2. Loop detection function
A. Simultaneously supply 24V loop and measure current.
B. Built-in 250 ? HART loop resistor for easy access to HART communication devices without the need to carry load resistors separately.
3. Analog output function
A. 4~20mA current setting by bit, manual step output function.
B. SIMULATE function simulates the transmitter.
C. Can be displayed in mA and percentage.
D. The output accuracy is 0.2%, and the output is displayed as 5 bits.
4. Measurement function
A. AC voltage, DC voltage, ohmic, DC current, on and off, diode.
B. The measurement accuracy is 0.5%, and the current measurement is displayed as 5 bits, which is displayed as 4 bits.
C. Measurement rate: 2.5 times / sec
D. Can be displayed in mA and percentage.
E. Data retention function.
5. Use panel calibration technology.
Specifications:
Specifications:
Display Screen: display in four digits(5 in current measurement and output)
Refresh display:2.5 displays/second
Operation temperature and moisture range :0~40 ºC, relative moisture≤85%(no condensation)
Storage temperature and moisture range:-20 ºC~60 ºC,relative moisture≤90%( no condensation)
Accuracy required temperature and moisture range:23±5ºC,relative moisture≤75%( no condensation)
Temperature co-efficiency:0.1 × basic accuracy / ??temperature range <18??or >28??
Continuity/open test:Beeper sounds for resistance readings below threshold, or to indicate a momentary open circuit
Battery type:two AA 1.5V Alkaline (LR6) batteries(Not included)
DC current output(SIMULATE):about 50 hours;
DC current output (SOURCE)20mA(1000Ω load):about 2.5 hours.
Battery life:when using alkaline battery:measures any parameter: about 100 hours;
Low battery:displays battery symbol
Automatic power-off:The default value is 5 mins if no operation which is adjustable.
Preheating time:10min
Measuring detailed accuracy indicator
Outputting detailed accuracy indicator
Package inlcudes:
1 x Multimeter
Detail pictures:
Measuring detailed accuracy indicator
Measuring part | range | frequency | accuracy | remark |
DC voltage | 50V | 10mV | 0.5%+4 | • Measuring impedance 10 MΩ (standard value) • Common mode restrain: 50Hz or 60Hz > 100Db • String mode restrain:50Hz or 60Hz > 45dB • Overload protection: 600V (pea'k –pea'k value) |
AC voltage | 500V | 100mV | 0.5%+20 ( 45Hz~100Hz ) | • frequency beeps: 45Hz~400Hz • Adaptive to range from 10% to 100% • AC conversion: mean value • Measuring impedence:10 MΩ(standard value)<100pF • Common mode restrain: 50Hz or 60Hz > 100Db • Overload protection: 600V(pea'k –pea'k value) |
2%+20 ( 100Hz~400Hz ) | ||||
Ohm | 5kΩ | 0.001kΩ | 0.5%+4 | • open-circuit voltage: <5V; • short-circuit current: about 0.1mA. • leads resistance excluded in the accuracy; • overload protection: 600V |
continuity | 500Ω | 0.1Ω | Short-circuit beep is about 20Ω | • open-circuit voltage: <5V • short-circuit current: about 0.1mA. • overload protection: 600V(pea'k –pea'k value) |
diode |
2V | 0.001V | 1%+20 | |
DC current | 20mA | 0.001mA | 0.2%+4 | • overload protection: 100mA/250V fast-melt protection; • loaded voltage: about 18mV/mA |
Outputting detailed accuracy indicator
output function | range | output designated range | frequency | accuracy | remark |
DCI DC current |
20mA | 0.000~22.000mA | 0.001mA | 0.2%+4 | • Max. load: 1kΩ at 20mA |
SIMULATE Simulate transmitter |
-20mA | 0.000~-22.000mA | 0.001mA | • external power supply: 5~28V • Max. load: 1kΩ at 20mA |
|
LOOP Loop-circuit power |
24V | ±10% | • Max. output current 25mA |
Package inlcudes:
1 x Multimeter
Detail pictures: